SKU : OPTIZEN™ Alpha
The luxuriously designed OPTIZEN Alpha is spectrophotometer to use double-beam method.
หมวดหมู่ : 12. Lab Equipment ,  Spectrophotometer ,  KLAB (Korea) , 
Share
OPTIZEN Alpha can grasp the quantitative characteristics such as density or purity by measuring transmittance or optical density according to wavelength of sample in the range of ultraviolet rays and visible ray.
OPTIZEN SERIES can be utilized from not only a general analysis experiment to but also a specialized research field and guarantee accurate measurement and excellent reproducibility, accordingly offer reliable results in the various fields such as environment, biotechnology, chemistry, etc.
Product Qualification
Photometrics System | Double-beam type |
---|---|
Monochromator | Czerny-Turner type with 1200 lines/nm blazed grating |
Light Source | Tungsten Halogen Lamp & Deuterium Lamp |
Lamp interchange Wavelength | 340~410 nm (Default 370 nm) |
Detector | Silicon Photodiodes |
Spectral Bandwidth | 1 nm (190 to 1100 nm) |
Wavelength Range | 190 to 1100 nm |
Wavelength Setting | 0.05 nm |
Wavelength Accuracy | ±0.3 nm (For entire range) ±0.1 nm (656.1 nm) |
Wavelength Repeatability | <± 0.1 nm |
Slew Rate | About 8,800 nm/min |
Scan Speed | max 6,000 nm/min |
Photometric Range | Absorbance : -4 to 4 Abs Transmittance : 0 % to 400 % |
Photometric Accuracy | ±0.002 Abs at 0.5 Abs ±0.004 Abs at 1.0 Abs ±0.006Abs at 2.0Abs |
Photometric repeatability | ± 0.0002 at 0.5 Abs ± 0.0006 at 1.0 Abs ± 0.001 at 2.0 Abs |
Baseline stability | < 0.0003 Abs/h |
Baseline flatness | <± 0.0005 Abs |
Stray light | < 0.02 % Nal at 220 nm, NaNO₂ at 340 nm |
Standard cell Holder | Autometic Rotary type 8-position Multi-Cell Holder |
Operating system(OS) | Windows 10 (Embedded PC); |
Display | 8-inch color LCD with touch screen |
Power Requirement | 100~240 V; 50~60 Hz |
Dimensions(WxDxH) | 520(W)x500(D)x200(H) |
Weight(kg) | 14 kg |
PC Software | (optional) OPTIZEN View for Windows® (optional) OPTIZEN Secure™ |
* The contents of this document may be changed without notice.
Product details
The previous single beam type spectrophotometer has a weak point of error occurrence in measuring a sample, because of the movement of a light source by time lag in measuring the strength of between a reference light and a light from sample.
To solve the problem, OPTIZEN Alpha was designed as a double-beam type spectrophotometer.
The system utilizes an additional reference beam to improve the measurement performance by compensating the intensity fluctuation of its light source.
Product Information