Autoclave, Class B, Benchtop Type, STB-B-2B Series
Description
Application:
Suitable to sterilizing wrap, unwrap, solid, porous, hollow device. Suitable for the Dept of Stomatology, Ophthalmology, Surgery and Lab.
Specifications
Model classification
Model | Power | Voltage | Frequency | Tank Capacity | Appearance size (mm) | Inside Dimension (mm) | Package Dimension (mm) | G·W |
STB-B23-2B | 2000 W | AC220V/110V | 50/60Hz | 23Litres | 445×640×395 | Φ249×450 | 740×550×500 | 53 kg |
STB-B18-2B | 1800 W | AC220V/110V | 50/60Hz | 18Litres | 445×550×395 | Φ249×355 | 670×550×500 | 48 kg |
STB-B16-2B | 1800 W | AC220V/110V | 50/60Hz | 16Litres | 445×550×395 | Φ230×360 | 670×550×500 | 47 kg |
STB-B12-2B | 1800 W | AC220V/110V | 50/60Hz | 12Litres | 445×550×395 | Φ200×360 | 670×550×500 | 46 kg |
Technical parameter
Sterilization level | CLASS B (European Class B) |
Sterilization temperature | 134/121 |
Display accuracy | ±0.5 |
Recording accuracy | ±0.5 |
Sensor accuracy | 0.01 |
Special sterilization | Kill active HV,HBV,BSE and spore |
Drying program | Strong vacuum drying,the residual humidity<0.2% |
Display | Widescreen LCD display |
Test | Bowie & Dick steam penetration test, Vacuum test, Helix spiral test |
Safety | Safety valve, pressure lock, fault self-check |
Sterilization record | Optional USB, printer (built-in or external) to record the sterilization process and date |
Water supply system | 3L distiller water tank,3L waste water tank |
Cleaning program | Automatic clean inside pipe and steam generator |
Sterilization configuration | 5 layers, 3 trays (circular shape, more space in actual use) |