The low-temperature centrifuge is widely used in various fields such as biology, medicine, agriculture, and chemical industry, lt is an instrument that utilizes centrifugal force to perform gravity-based fractionation of samples. The maximum speed of this machine is n=12000rpm, and the centrifugal force is RCF=13684xg. This low-temperature centrifuge ensures that the samples do not heat up during centrifugation, providing optimal protection for the samples.
Product Features
True low temperature, centrifuge without heating up
Truly protect samples, reduce RNA/protein degradation, and achieve better experimental results
1. Unsymmetrical balancing monitoring system: if the sample is unsymmetrically placed and unbalanced, the instrument will automatically stop centrifuging.
2. Compatible with various rotors: suitable for 1.5/2mL (standard), 5mL and 10mL centrifuge tubes.
Working Environment
Power Parameters | 200-240VAC, 50-60Hz, 700W; power voltage fluctuation does not exceed 10% of the rated power voltage. |
Room Temperature | 5-40 |
Relative Humidity | Maximum 80% |
Operating Location | Indoor use |
Altitude | Up to 2000 meters |
Technical Parameters
Dimensions | 530×355×268 mm |
Weight | 45 kg |
Screen Size | 5 inches |
Temperature Control Accuracy | ±2 |
Operating Time | 1-99 minutes |
Operation Parameters
Operating Mode | Continuous |
Time Setting | 1-99 min |
Balancing Monitoring | Gyroscope |
Refrigeration system | SECOP Compressor Cooling |
Temperature Range | -20 - Room Temperature |
Temperature Efficiency | Specimens at room temperature go in (centrifuge for 10 minutes) and come out at 4 |
Rotor Capacity | [1.5mL / 2mLx24] [5mLX12] [10mLX6] |
Rotor Replacement | Quick-change mode: Only requires a dedicated wrench for removal and installation |
Rotor Removal | Can be removed as a whole without opening the lid, to prevent the dispersion of specimen gas molecules |
Specimen safety | Equipped with multiple sealing rings to prevent dispersion contamination |
Interface Operation | Full touch screen interface, can be operated with gloves |
Machine Safety | Double lock mechanism, and equipped with manual door opening mechanism during power failure |
Status Indicator | LED strip status indicator, displays real-time working status |
Software Functions | [Quick cooling setting] [Real-time temperature monitoring] [User program memory (up to 10 sets)] [Real-time parameter adjustment] [Gyroscope balancing detection] [Real-time monitoring of motor overvoltage, undervoltage, blockage, and overheating, with alarms and automatic shutdown] |
Product Structure
Operation Instruction
1. Unpacking and Installation
1) Remove the low-temperature centrifuge from the packaging box and place it in a suitable location, ensuring all components are complete;
2) Place the machine on a flat and level experimental table to prevent vibration during operation;
3) Let it stand for 10 hours to ensure the coolant is in place, connect the power supply (200-240V), and turn on the power switch at the back of the low-temperature centrifuge;
4) Turn on the instrument power switch, wait for 10-15 minutes, and observe if the temperature inside the centrifuge chamber reaches the preset temperature;
5) If the temperature reaches the preset temperature, the machine is ready for use
2. Operation Interface
1) Open door button,Opens the centrifuge lid
2) Start/Stop button.Starts and stops centrifugal separation
3) Set speed.The speed can be displayed in terms of centrifugal force (xg) or revolutions per minute (rpm). The current centrifuge is set to display rpm and xg. Click on the centrifugal force (or speed value) to modify the parameters, input the desired value for one, and the other will automatically adjust accordingly.
4) .Set time.Input the required parameter values in minutes (range: 1-99) and seconds (range: 0-59)
5) .Instant release button.Instant centrifugal separation.Press and hold the button to start centrifugation, release to stop centrifugation
6) Set temperatureAdjust the temperature parametersRange: -20 to room temperature
7) Speed up/slow downRange: 1-10 (the larger the number, the slower the speed)
Gear | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |
Acceleration Time/s | 20 | 25 | 30 | 35 | 40 | 45 | 50 | 55 | 60 | 70 |
Deceleration Time/s | 20 | 25 | 30 | 35 | 40 | 45 | 50 | 60 | 70 | 80 |
8) Parameters.View 10 sets of parameter values in the background
9) Rapid cooling button.Click to enter rapid cooling mode
3. Operation Steps
1) Plug the power cord of the low-temperature centrifuge into a 220V power outlet with grounding protection and turn on the power switch at the back of the low-temperature centrifuge;
2) After the startup interface is displayed, wait for about 5 seconds to enter the main operation interface;
3) After the low-temperature centrifuge is turned on, the refrigeration system starts to operate. Place the centrifuge tubes symmetrically into the rotor and cover the rotor lid. Then close the top cover of the centrifuge machine;
4) Set the centrifugal force (or speed value), time, temperature, acceleration, and deceleration;
5) Click the button to start the centrifugation process. The instrument will automatically stop after the countdown reaches zero.
4. Parameter Settings
The instrument has 1 set of preset parameters, with a total of 10 sets of parameters, each of which can be edited.
For example, click on to enter the parameter setting interface, select the centrifugal force (or speed value), acceleration, deceleration, and time for modification. After the modification is completed, click "Save" to save the corresponding parameters.
Click on button to start the device. The touchscreen will countdown according to the set time. After the countdown is complete, the centrifugation ends, and the instrument automatically stops. Wait for the low-temperature centrifuge to completely stop before opening the top cover, unscrew the fixing screws, and remove the adapter. If the centrifugation is not complete, repeat the above steps until the centrifugation requirements are met
5. Alarm Prompt
When you click start, if the instrument detects an unbalanced rotor, this interface will pop up and the experiment will be stopped.
Once this interface appears, there will be continuous alarm sounds until you click "OK" to dismiss it.
6. Breathing Light Prompt Comparison Table
No. | Instrument Status | Breathing Light Color | Breathing Light State |
1 | Self-test | Magenta | Fast breathing |
2 | Fault | Red | Steady on |
3 | Standby | Green | Breathing |
4 | Door not closed properly | Yellow | Fast breathing |
5 | Acceleration | Blue | Flowing |
6 | Deceleration | Blue | Flowing |
7 | Arrival at specified speed | Blue | Steady on |
8 | Rapid cooling | Cyan | Flowing |
9 | Rapid cooling reaches standard | Cyan | Steady on |
10 | Instant centrifugation | White | Flowing |
Comparison of Rotation Speed and Centrifugal Force
n(rpm) | RCF(g) |
1000 | 95.03 |
2000 | 380.12 |
3000 | 855.27 |
4000 | 1520.48 |
5000 | 2375.75 |
6000 | 3421.08 |
7000 | 4656.47 |
8000 | 6081.92 |
9000 | 7697.43 |
10000 | 9503 |
11000 | 11498.63 |
12000 | 13684.32 |
13000 | 16060.07 |
14000 | 18625.88 |
15000 | 21381.75 |
16000 | 24327.68 |
17000 | 27463.67 |
18000 | 30789.72 |
19000 | 34305.83 |
20000 | 38012 |
Equipment Maintenance
1. If solvents or salt water, acidic or alkaline solutions are spilled on the low-temperature centrifuge, use a soft cloth dampened with water to wipe them clean immediately to avoid damage to the instrument.
2. Make sure to turn off the power and unplug the power cord before cleaning.
3. Do not perform high-pressure sterilization on any part of the low-temperature centrifuge, including the adapter.
4. The components of the adapter should be cleaned after use. Use appropriate cleaning agents, rinse with distilled water, and then dry with paper towels.
5. After using suitable cleaning agents, wipe the low-temperature centrifuge clean with a soft cloth.
6. Recommended disinfectants and cleaning agents for cleaning the low-temperature centrifuge and adapter include:
a. General cleaning agent
b. Neutral detergent
c. 70% alcohol
Standard Adapter
(2 mL × 24)